Multi-channel independently controlled split Syringe Pump

Multi-channel independently controlled split Syringe Pump

No. of Syringes:1/2/3/4/5/6/7/8

Flow rates:0.0008uL/min - 108.178mL/min

Display:Liquid volume, flow rate, or linear velocity

Control:5-inch touchscreen, external control

Stroke of the drive unit:90mm

Motor:Stepper motor

No. of Syringes:1/2/3/4/5/6/7/8
Flow rates:0.0008uL/min - 108.178mL/min
Display:Liquid volume, flow rate, or linear velocity
Control:5-inch touchscreen, external control
Stroke of the drive unit:90mm
Motor:Stepper motor

  • Product Parameters
  • Selection Guide
  • Service Guarantee
  • Related Products
  •  
    Technical Specifications
    Number of Channels 1 / 2 / 3 / 4 / 5 / 6 / 7 / 8
    Running Control Each drive unit can be controlled separately
    Working Mode Infusion / Withdraw ( Infusion, withdrawal, infusion / withdrawal, withdrawal / infusion, continuous )
    Max. Infusion Distance 90 mm
    Syringe Size 10 uL to 60 mL
    Syringe Selection The syringe can be selected in the manufacturer table which includes manufacturer, material and size or input the inner diameter of the syringe barrel directly
    Display Iindustrial grade color touch screen
    Setting Mode Touch screen control
    Fast forward / backward Infuse or withdraw liquid at full speed
    Linear Speed Min: 3um/min   Max: 90 mm/min
    Adjusting Resolution 3um
    Linear Force > 90N
    Travel resolution 0.3125 um/ustep
    Travel control accuracy ≤ 0.5%
    External control signal RS485 or Start/Stop
    Communication Interface RS485
    Power Supply (Controller) DC 24 V, 120W
    Operating temperature 0°C - 40°C
    Relative humidity < 80%
    Controller Dimensions 245mm*188mm*102.28mm
    Controller Weight 1.8 Kg
    Drive Unit Dimensions 259mm*99.4mm*133mm
    Drive Unit Weight 2.4 Kg
    IP rating IP31
    Power-off Memory Storing the running parameters automatically
    Calibration Function Acquire accurate volume through calibration
    Protection Function The pump will stall and give an alarm when the drive structure of the pump is blocked
    Syringe Protection Adjust syringe rest to prevent syringe from damaging
     
    Syringe Reference traffic
    Specification Inner diameter(mm) uL - mL/min
    10uL 0.485 0.001-0.024
    25uL 0.729 0.001-0.054
    50uL 1.03 0.002-0.108
    100uL 1.457 0.005-0.217
    250uL 2.304 0.013-0.542
    500uL 3.256 0.025-1.082
    1mL 4.699 0.052-2.254
    2.5mL 4.851 0.055-2.403
    3mL 8.585 0.174-7.525
    5mL 11.989 0.339-14.676
    8mL 9.525 0.214-9.263
    10mL 14.427 0.490-21.251
    20mL 19.05 0.855-37.053
    30mL 21.59 1.098-47.592
    50mL 26.594 1.666-72.211
    60mL 29.45 2.044-88.553


     

  •   Selection Strategy

      1. Confirm the required flow rate (mL/min) and the number of pipelines to be conveyed.

      2. Confirm the usage scenario – it can be used directly or in conjunction with equipment, controlled via RS485 or external signals.

      3. Parameter display – digital tubes only show speed; LCD screens display flow rate, speed, and other calibration parameters; touch screens display flow rate, speed, and other programming or configuration parameters.

      4. Liquid composition – affects the choice of tubing. For conventional fluids, platinum-cured silicone tubing is recommended; for corrosive fluids, imported tubing is recommended.
     

      Considerations when selecting a driver:

      - Flow rate – the higher the driver speed, the greater the flow; select an appropriate driver based on the reference flow rate.

      - Control method – options include fixed speed, adjustable speed, adjustable speed with flow correction (for precise flow control), adjustable speed with flow distribution (quantitative distribution), and external signal control, among others.

      - Parameter display – choices include digital tubes, LCD screens, and touch screens.

      - Protection – depending on the usage environment, consider the sealing, dust-proof, and waterproof requirements for the driver.
     

      Considerations when selecting a pump head:

      - Flow rate – select an appropriate pump head according to the reference flow rate.

      - Tubing replacement frequency – except for OEM series products, most pump heads meet the requirements for quick tubing replacement.

      - Suitable number of channels – for multi-channel requirements, you can choose a multi-channel pump head or connect pump heads in series.

      - Chemical resistance – consider the impact of fluid overflow on pump head materials.
     

      Considerations when selecting tubing:

      - Flow rate – refer to the tubing’s reference flow rate curve.

      - Chemical resistance – select the material of the tubing accordingly.

      - Fluid temperature – different materials perform differently at various temperatures.

      - Suction requirements – high-performance tubing should be selected if suction requirements exist.

      - Tubing lifespan – the lifespan varies depending on usage conditions; for example, BPT tubing is nearly 30 times more durable than domestic silicone tubing.

     


  •   Regarding Delivery Time

      Peristaltic pumps, syringe pumps, micro plunger pumps, domestically produced silicone tubing, and other series of products are mostly in stock and can be shipped within 48 hours;

      Filling machines, filling systems, and other products are usually made to order, with relatively low inventory. The delivery time is about 1–2 weeks, with a small portion in stock;

      Common models of imported hose series products are stocked; due to the impact of international logistics on imported hoses, under special circumstances, the delivery time is about 1–4 weeks;

      Note: All peristaltic pump series products require 3 days of continuous operation and aging before leaving the factory to ensure a lower failure rate and allow customers to use them with confidence.
     

      Regarding After-Sales

      Warranty is one year (excluding consumables; improper use or human damage is not covered)

      For special situations caused by selection or actual application, such as flow differences, usage environment, or personal reasons, products can be directly exchanged by contacting customer service, as long as it does not affect secondary sales.
     

      Regarding Logistics

      Please communicate with customer service and confirm.